9 Repair
9.2 Height adjustment motor (M1_4)/spindle
370
9.2

Height adjustment motor (M1_4)/spindle

9.2.1

Preparing for motor replacement

1. Switch the unit on.
2. Use the Up/Down keys on the control panel to move the slide up.
3. Switch the unit off again.
4. Remove the covers:
– Intermediate piece
– Profile covers (top and bottom)
Tip: While loosening the screws, press the top profile cover down
towards the unit and allow it to slide down once the screws are loose.
– Arm cover
– Slide cover rear, center
– Slide cover rear, top
– Slide cover rear, bottom and
– Slide cover front.
Tip: If the height adjustment motor is inoperative, you can also move the
slide manually [ → 370].
9.2.1.1

Moving the slide manually

CAUTION
Risk of injury due to uncontrolled movement of the slide
If the slide can no longer be moved electrically, it must be moved
mechanically.
The position of the slide must be secured to ensure that no uncontrolled
downward movement occurs during service, in cases where the
carriage has fewer self-locking properties.
For this purpose, Sirona recommends using the height adjustment
service kit,
REF. 62 57 518. This service kit is used to prevent automatic downward
movement of the slide during service by fixing the slide and the spindle
holder.
The clamp (A) should be clamped under the slide. The locking pin (B) is
used to secure the spindle holder against twisting.
It must be ensured that no one is located underneath the ring arm during
the repair.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016