Checking The Data Paths; Creating 2D Test Images - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

6 Troubleshooting
6.9 Checking the data paths
180
6.9

Checking the data paths

The 2D and 3D data paths can be checked by creating panoramic, ceph
and sensor (3D) test images. To create test images, proceed as follows:
SIDEXIS XG must be open.
➢ In SIDEXIS XG, call the
"Utilities"
/
"Service exposure "
test images"
6.9.1

Creating 2D test images

"Test exposures/test images"
The menu
1. In the structure tree, under
image"
element.
The menu for creating the panoramic and ceph test image
appears in the action window.
2. Click on the
Ceph"
button to create the appropriate test image.
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
3. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
The test image appears in the exposure window.
Service Manual ORTHOPHOS XG 3D / Ceph
"Test exposures/test images"
"Constancy test..."
"3D"
/
"Select X-ray component"
/
"Test exposures 2D"
"Image acquisition Panorama"
Sirona Dental Systems GmbH
menu:
"Select X-ray device"
/
/
"Test exposures/
/
is called [ → 469].
"2D test
, select the
"Image acquisition
or
63 03 510 D3352
D3352.076.03.11.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents