2 Safety instructions
2.6 Condensation
16
2.6

Condensation

Extreme fluctuations of temperature may cause condensation inside the
Safety information for condensation: Service engineer
unit. Do not switch the unit on before it has reached normal room
temperature. See also the chapter "Technical details"
2.7

Laser light localizer

The system incorporates Class 1 laser products.
Safety information for light localizer: Service engineer
A minimum distance of 10 cm (4") is required between the eye and the
laser. Do not stare into the beam.
Do not use the system with any other lasers, and do not make any
changes to settings or processes that are not described in these
operating instructions. This may lead to a dangerous exposure to
radiation.
2.8

Ventilation slots

Never cover the ventilation slots on the unit under any circumstances,
Ventilation slots
since this may obstruct air circulation. This can cause the unit to overheat.
2.9

Qualifications of service personnel

Installation and startup may be carried out only by personnel specifically
authorized by Sirona.
2.10

Radiation protection

The valid radiation protection regulations and measures must be
Safety information for radiation protection: Service engineer
observed. The statutory radiation protection equipment must be used.
During an exposure, the service engineer should move as far away from
the X-ray tube assembly as the coiled cable of the manual release
permits.
With the exception of the service engineer, no other persons are allowed
to stay in the room during an exposure.
In case of malfunctions, cancel the exposure immediately by letting go of
the exposure release button.
NOTICE
Information about radiation protection for Canada
3D imaging should not be used for screening examinations. 3D imaging
examinations must be clinically warranted and each exam must be
justified by demonstrating that the benefits outweigh the risks.
NOTICE
Where it is likely that evaluation of soft tissues will be required as part of
the patient's radiological assessment, the appropriate imaging should
be conventional medical CT or MR, rather than 3D imaging using Cone
Beam technology.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016