9.3.3

Laying of cables when replacing the ring motor

Laying cables for ring motor replacement: XG 3D
1. Lay the cable (G) in a loop from top to bottom and fix it in the cable
clamp (J).
2. Plug connector X813 (H) into board DX1.
9.3.4
What has to be done after replacing the ring motor
(M1_3)/pinion?
1. Check the function of the ring motor.
2. Perform complete unit adjustment or calibration [ → 184].
9 Repair
9.3 Ring motor (M1_3)
383