What has to be done after replacing the user interface?

9.6.1.1
IMPORTANT: So that the board is also replaced with the user interface,
Important: XG 3D cross reference
you MUST also follow the instructions in the chapter titled "Measures
following replacement of boards [ → 432]".
1. Check that the user interface and the display elements are
functioning correctly: When the unit is switched on, all of the display
elements must light up briefly.
2. Perform a software update to the latest version [ → 62].
Following replacement of the user interface, the language set on board
DX7 is set to the factory setting by default (00 = German, English, French,
Italian). If the configured unit language set (which can be queried by
running service routine S017.5 or using the "extended detail query" in
SiXABCon) has a configuration other than 00, this configuration will be
copied to board DX7 by the update function.
9 Repair
9.6 Control panel