Error Messages; Error Code: Ex Yy Zz; Ex - Error Type - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
x
Description
1
System warning
System message
2
Errors caused by
system overload
3
The system detects that
a key was pressed
during power-on.
63 03 510 D3352
D3352.076.03.11.02 09.2016
5.4

Error messages

Error messages are displayed as error codes (Ex yy zz) on the Easypad
XG 3D error messages
touchscreen and on the remote control display (if there is one).
The codes provide you with error type, error location and troubleshooting
information.
5.4.1

Error code: Ex yy zz

The error messages are encoded according to the following pattern:
Ex
Error type
yy
Location
zz
Consecutive
number
5.4.2

Ex - Error type

Identifier x is intended to help you reach a decision quickly on how to
proceed with the corresponding error.
Error group
This error group includes all errors
that indicate still acceptable
tolerance variations, or messages
about states which do not directly
affect system operation.
This error group includes states
that indicate temporary
overtemperatures or similar, for
example. The cause of the error
disappears automatically after a
certain waiting time.
This error group includes all errors
that indicate invalid signal states
of keys and safety signals during
power-on.
"Troubleshooting" classification for the user
Module, subsystem or logical function unit
Identification of error
Actions required
● Acknowledge the error message.
If the error occurs again ...
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure
If the error occurs again ...
● Run an error diagnosis [ → 94].
● Acknowledge the error message.
● Repeat the procedure step after a certain
waiting time.
If the error occurs again ...
● Extend the waiting time.
If the error occurs again ...
● Run an error diagnosis [ → 94].
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure
If the error occurs again...
● Run an error diagnosis [ → 94].