Checking The Radiation Time - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
D
Preparing the measurement
A
63 03 510 D3352
D3352.076.03.11.02 09.2016
4. Reattach the cover plate (G) to the tube assembly with the 4 screws
(F).
F
G
5. Reattach the lid of the tube assembly cover (E) to the unit and secure
D
it with the 3 screws (D).
D
E
10.10.3

Checking the radiation time

1. Attach the Mult-O-Meter sensor in the middle of the combisensor (on
2D sensor side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
The unit moves to its starting position.
10 Maintenance
10.10 Checking the tube data
467