Checking The Ceph Sensor Holder - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

10 Maintenance
10.7 Checking the ceph sensor holder
Check whether the sensor locks and
unlocks easily and is firmly seated in the
holder
452
10.7

Checking the ceph sensor holder

1. Carefully insert the sensor upward into the holder: The sensor must
snap in place audibly.
A
2. Check the sensor for firm seating.
3. Press the locking button and carefully pull the sensor downward out
of the holder: The sensor must easily be removed.
If the sensor cannot easily be inserted in or removed from the holder,
or is seated too loosely in the holder:
Check the magnets on the sensor for dirt or foreign particles and
B
clean them or remove any foreign particles if necessary.
Replace the sensor holder [ → 424].
Replace the sensor (see Operating Instructions).
C
Sirona Dental Systems GmbH