Run Procedure - Beckman Coulter JA-25.50 Manual

Fixed angle rotor
Hide thumbs Also See for JA-25.50:
Table of Contents

Advertisement

JA-25.50 Rotor

RUN PROCEDURE

10
Microprocessor-Controlled J2 Series Centrifuges
(Models J2-MI and J2-MC)
1. Press the [
] key on the centrifuge control panel and then use
TEMP
the keypad to enter the sample temperature.
2. Find the compensation value in Table 1 that corresponds with the
set temperature and run speed.
3. Press [
]. The word "COMP" flashes below the
COMP ADJ
TEMPERATURE display and the display flashes.
4. Use the keypad to enter the required compensation value. Press
the [
] key to enter a minus sign; pressing it again will remove the
±.
minus sign.
5. Check the temperature display. (If the entry is incorrect, press [
and reenter the digits.)
6. When the entry is correct, press [
NOTE
To clear a COMP ADJ entry, press
[
], [
COMP ADJ
When the rotor is properly loaded and installed in the centrifuge, you
are ready to perform the run. Refer to the applicable centrifuge
instruction manual for operating instructions.
!
WARNING
Operator error or tube failure may generate
aerosols. Toxic, pathogenic, or other
hazardous materials must not be run in
this rotor unless all appropriate safety
precautions are taken. Also see the Safety
Notice at the beginning of this manual.