RUN PROCEDURE

Microprocessor-Controlled J2 Series Centrifuges (Models J2-MI,
J2-21M, J2-MC, and J2-21M/E)
Enter rotor code 14, then follow the steps below.
1. Press the [
] key on the centrifuge control panel and then use
TEMP
the keypad to enter the sample temperature.
2. Find the compensation value in Table 1 that corresponds with the
set temperature and run speed.
3. Press [
]. The word "COMP" flashes below the
COMP ADJ
TEMPERATURE display and the display flashes.
4. Use the keypad to enter the desired compensation value. Press the
[
] key to enter a minus sign; pressing it again will remove the
±.
minus sign.
5. Check the temperature display. (If the entry is incorrect, press [
and reenter the digits.)
6. When the entry is correct, press [
NOTE
To clear a COMP ADJ entry, press [
[
], and [
0
ENTER/RECALL
!
WARNING
Operator error or tube failure may generate
aerosols. Toxic, pathogenic, or other
hazardous materials must not be run in
this rotor unless all appropriate safety
precautions are taken. Also see the Safety
Notice at the beginning of this manual.
J-Lite JLA-16.250 Rotor
].
ENTER/RECALL
],
COMP ADJ
].
]
CE
13