Cleaning The 4 L Waste Container; Cleaning - Beckman Coulter CytoFLEX Series Instructions For Use Manual

Flow cytometers
Table of Contents

Advertisement

3
Remove the sheath fluid harness from the sheath fluid container.
4
Empty the residual sheath fluid from the sheath fluid container.
5
Add about 50 to 100 mL of CytoFLEX Sheath Fluid to the sheath fluid container.
6
Insert the sheath fluid harness back into the sheath fluid container and tightly close the sheath
fluid container cap.
7
Swirl the sheath fluid in the sheath fluid container, rinsing all surfaces.
8
Empty the sheath fluid container.
9
Refill the sheath fluid container. Refer to
CHAPTER 11, Replacement/Adjustment

Cleaning the 4 L Waste Container

Clean the waste container once a month.
1
Confirm that the instrument is turned off or is in the standby state.
2
Remove the waste container from the Fluid Container holder.
WARNING
Risk of biohazardous contamination if you have skin contact with the waste
container, its contents, and its associated tubing. The waste container and its
associated tubing could contain residual biological material and must be handled
with care. Clean up spills immediately.
3
Remove the harness from the waste container.
B49006AH
Filling the 4 L Sheath Fluid Container [CytoFLEX]
Procedures.
Cleaning Procedures
10
Routine Cleaning
in
10-11