System Overview
Fluidics System
Figure 1.8 Sheath Tank
b
1. Sheath outlet. Connects the sheath tubing to the sheath filter.
2. Air inlet. Provides pressure to convey the sheath to the flow cell or wash station.
3. Safety valve. Releases the pressure from the tank in case of emergency.
4. Pressure relief valve. Releases the pressure from the tank manually.
5. Sheath tank scale. Monitors the volume of sheath remaining in the sheath tank.
NOTE
tank scale first. Refer to
Replacement/Adjustment

Shutdown Fluid Container

The shutdown fluid container is a polypropylene container with 1-L capacity. Refer to
accommodates the shutdown fluid for decontaminating the fluid lines. The container does not
require pressurization.
1-10
If you need to switch between the 4 L sheath tank and the 7 L sheath tank, reset the sheath
Resetting the Sheath Tank Scale
Procedures.
c
d
e
f
in
CHAPTER 11,
Figure
1.9. It
C37808AC