Installing new or upgrading existing software

It is possible to install new software or new software components.
Different software (components) can be installed:
• language files,
• modality specific files,
• LUT,
• calibration files,
• ...
The new software will be activated after the installation procedure (this may
require a reset operation). The previously installed software components will
be automatically removed when installing new software or new data. A copy
of the previous configuration is saved (as a dataset) on the internal media.
1. Press the Key-operator key to enter the Key-operator mode.
2. On the Key-operator main menu, press the Down key eight times, followed
by the Confirm key to select 'Installation'.
The 'Installation' screen will be displayed:
3. Press the Confirm key to select 'Installation from USB-stick'.
The following screen is displayed:
• If you want to back up the current configuration on an external USB
flash drive, select 'Ok, make backup', followed by the Confirm key. The
printer will perform the 'General save configuration procedure'. Note:
the complete configuration is saved!
• If you do not want to back up the current configuration, press the Up/
Down arrow keys to select 'SKIP backup', followed by the Confirm key.
4. After the back-up is finished (or if you have chosen to skip the back-up),
the system will proceed with the actual upgrade.
The 'Installation from USB-stick' screen is displayed:
Drystar AXYS XL | Advanced Operation (key-operator mode) | 71
Note: Before the procedure starts, the system suggests you
make a back-up copy of the current configuration. It is
highly recommended to do this, so you can always restore it
in case something should go wrong.
3753B EN 20210520 0952