Remarks
•
After it is enabled, the instrument outputs a pulse with the specified edge
:OUTPut[<n>]:TRIGger:SLOPe
(
the beginning of the burst or sweep.
•
The trigger output is disabled when the burst trigger source is set to External or
the burst type is set to Gated. The rear-panel [AUX IN/OUT] connector cannot
be used for trigger input and trigger output at the same time.
•
The trigger output is disabled when the sweep trigger source is set to External.
•
When [<n>] is omitted, it is interpreted as CH1.
Return Format
The query returns 0 or 1.
Examples
:OUTPut:TRIGger ON /*Enables the trigger output for CH1.*/
:OUTPut:TRIGger? /*Queries whether the trigger output is enabled
for CH1. The query returns 1.*/
3.10.10
:OUTPut[<n>]:TRIGger:SLOPe
Syntax
:OUTPut[<
:OUTPut[<
Description
Sets or queries the edge of the trigger output signal for the specified channel.
Parameter
Name
<n>
<type>
Remarks
•
POSitive: outputs a leading edge pulse; NEGative: outputs a trailing edge pulse.
•
After the trigger output is enabled (
outputs a pulse with the specified edge from the rear-panel [AUX IN/OUT]
connector at the beginning of the sweep or burst.
•
When [<n>] is omitted, it is interpreted as CH1.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
n >]:TRIGger:SLOPe < type >
n >]:TRIGger:SLOPe?
Type
Discrete
Discrete
) via the rear-panel [AUX IN/OUT] connector at
Range
{1|2}
{POSitive|NEGative}
:OUTPut[<n>]:TRIGger
DG800 Pro/DG900 Pro Programming
Command System
Default
1
POSitive
), the instrument
Guide
61