Output[<N>]:Load - Rigol DG800 Pro Series Programming Manual

Function/arbitrary waveform generator
Hide thumbs Also See for DG800 Pro Series:
Table of Contents

Advertisement

:OUTPut1:IDLE? /*Queries the idle level position of the burst
signal. The query returns TOP.*/
3.10.2
:OUTPut[<n>]:LOAD
Syntax
:OUTPut[<
:OUTPut[<
Description
Sets or queries the output impedance for the specified channel.
Parameter
Name
<n>
<ohms>
<lim_set>
<lim_query>
Remarks
When [<n>] is omitted, it is interpreted as CH1.
The parameter <ohms> indicates setting the output impedance to a specified
value. The <lim_set> allows you to set the impedance to HighZ (INFinity), the
minimum value (MINimum), the maximum value (MAXimum), and the default
value (DEFault).
The output impedance setting affects the output amplitude and DC offset. If the
actual load impedance differs from the value specified, the voltage level
displayed would not match the voltage level of the device under test. To ensure
correct voltage level, the load impedance setting must match the actual load.
Return Format
The query returns the output impedance in scientific notation. For example, the query
might return +1.000000000000000E+02, indicating that the output impedance is 100
Ω. If the output impedance of the output connector of the specified channel is set to
HighZ (INFinity), the query returns 9.9E+37.
Examples
:OUTPut1:LOAD INFinity /*Sets the output impedance to HighZ for CH1
output connector.*/
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
n >]:LOAD {< ohms >|< lim_set >}
n >]:LOAD? < lim_query >
Type
Range
Discrete
{1|2}
Integer
1 Ω to 10 kΩ
{INFinity|MINimum|MAXimum|
Discrete
DEFault}
Discrete
{MINimum|MAXimum}
Command System
Default
1
50 Ω
-
-
DG800 Pro/DG900 Pro Programming
Guide