Counter:average:clear; Counter:average[:State] - Rigol DG800 Pro Series Programming Manual

Function/arbitrary waveform generator
Hide thumbs Also See for DG800 Pro Series:
Table of Contents

Advertisement

3.2.2

:COUNter:AVERage:CLEar

Syntax
:COUNter:AVERage:CLEar
Description
Clears the statistical results for counter measurements.
Parameter
None.
Remarks
This command is valid only when the statistical function

:COUNter:AVERage[:STATe]

(
When the statistical function is disabled, the statistical results are automatically
cleared.
Return Format
None.
Examples
:COUNter:AVERage:CLEar /*Clears the statistical results for counter
measurements.*/
3.2.3
:COUNter:AVERage[:STATe]
Syntax
:COUNter:AVERage[:STATe] <
:COUNter:AVERage[:STATe]?
Description
Sets or queries the on/off status of the statistical function for the frequency counter.
Parameter
Name
<state>
Remarks
None.
Return Format
The query returns 0 or 1.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
) is enabled.
state >
Type
Range
Bool
{0|1|OFF|ON}
Command System
Default
0|OFF
DG800 Pro/DG900 Pro Programming
Guide
9