Source[<N>]]:Voltage:couple:mode - Rigol DG800 Pro Series Programming Manual

Function/arbitrary waveform generator
Hide thumbs Also See for DG800 Pro Series:
Table of Contents

Advertisement

Return Format
The query returns the amplitude in scientific notation. For example, the query might
return +2.500000000000000E+00, indicating that the output amplitude is 2.5 Vpp.
Examples
:SOURce1:VOLTage 2.5 /*Sets the output amplitude to 2.5 Vpp for
CH1.*/
:SOURce1:VOLTage? /*Queries the output amplitude for CH1. The
query returns +2.500000000000000E+00.*/
3.11.20.2
[:SOURce[<n>]]:VOLTage:COUPle:MODE
Syntax
[:SOURce[<
[:SOURce[<
Description
Sets or queries the type of the amplitude coupling mode for the specified channel.
Parameter
Name
<n>
<mode>
Remarks
The amplitude coupling mode can be set to the following:
-
-
When the coupling is enabled, if switching to ratio coupling from deviation
coupling causes either channel to exceed the amplitude upper/lower limit after
calculation according to coupling rules, the generator will automatically adjust
the parameters to avoid parameter overlimit. If the parameter overlimit is
caused by switching to deviation coupling from ratio coupling, the amplitude
coupling will be automatically disabled.
When [:SOURce[<n>]] or [<n>] is omitted, it is interpreted as CH1.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
n >]]:VOLTage:COUPle:MODE < mode >
n >]]:VOLTage:COUPle:MODE?
Type
Discrete
Discrete
OFFSet: amplitude deviation mode. You can use
[:SOURce[<n>]]:VOLTage:COUPle:OFFSet
the coupled channel.
RATio: amplitude ratio mode. You can use
[:SOURce[<n>]]:VOLTage:COUPle:RATio
coupled channel.
Range
{1|2}
{OFFSet|RATio}
to set the amplitude deviation of
to set the amplitude ratio of the
DG800 Pro/DG900 Pro Programming
Command System
Default
1
RATio
Guide
159

Advertisement

Table of Contents
loading

Table of Contents