Parameter
Name
<n>
<ratio>
Remarks
•
When modifying the ratio, if base channel parameters exceed the limits of the
coupled channel after coupling calculation, the generator will automatically
adjust the waveform phase to avoid parameter overlimit.
•
When [:SOURce[<n>]] or [<n>] is omitted, it is interpreted as CH1.
Return Format
The query returns the phase ratio in scientific notation. For example, the query
might return +2.000000000000000E+00, indicating that the ratio is 2.
Examples
:SOURce1:PHASe:COUPle:RATio 2 /*Sets the phase ratio of the phase
coupling to 2 for CH1.*/
:SOURce1:PHASe:COUPle:RATio? /*Queries the phase ratio of the
phase coupling for CH1. The query returns +2.000000000000000E+00.*/
3.11.12.5
[:SOURce[<n>]]:PHASe:COUPle[:STATe]
Syntax
[:SOURce[<
[:SOURce[<
Description
Sets or queries the on/off status of the phase coupling function for the specified
channel.
Parameter
Name
<n>
<bool>
Remarks
•
The phase coupling function is not available for DG821 Pro (unless the two-
channel upgrade option is installed).
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Type
Range
Discrete
{1|2}
Real
0.01 to 100
n >]]:PHASe:COUPle[:STATe] < bool >
n >]]:PHASe:COUPle[:STATe]?
Type
Range
Discrete
{1|2}
Bool
{0|1|OFF|ON}
Command System
Default
1
1
Default
1
0|OFF
DG800 Pro/DG900 Pro Programming
Guide
129