Flow Sensor Failure; Internal Function Failure - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.4.7

Flow Sensor Failure

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
7.4.8

Internal Function Failure

Property
Alarm text
Priority
Alarm condition
Reset criteria
Ventilator action
Action to take
108
Alarms
User manual
Description
Flow Sensor Failure
Medium
No data or erroneous data from the flow sensor
Correct data from the sensor is received again.
An information message remains until acknowledged by
the user.
The Vivo 2 will continue treatment but with the following
limitations:
Monitoring of leakage is disabled.
Volume measurements are disabled.
The patient cannot trigger breaths (applies to assisted
modes)
An information message about the limitations is displayed
on the screen.
Description
Int. Function Failure
High
Failure of internal function that prevents treatment or nor-
mal operation of the ventilator.
The error code that follows the alarm text indicates the kind
of function failure.
Correct function is restored.
The ventilator will stop the treatment.
Restart the Vivo 2.
If the alarm persists or reoccurs: Take a note of the error
code and contact your supplier of the Vivo 2 .
Doc. 007229 D-3