Ambient Pressure Compensation Lost Alarm; Low Battery Alarm; Critically Low Battery Alarm - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

7.4.5

Ambient Pressure Compensation Lost Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
7.4.6

Low Battery Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset Criteria
7.4.7

Critically Low Battery Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset Criteria
106 Alarms
Vivo 3 Clinician's Manual
Description
Pressure Comp Lost
Medium
An Ambient Pressure Compensation Lost alarm will be given
when the automatic ambient pressure compensation func-
tionality is out of order.
The Vivo 3 will continue treatment according to the current
settings.
Normal atmospheric pressure at sea level will be used as
approximation for the temporary ambient pressure compen-
sation. If used at other altitude, delivered and measured pres-
sures may deviate.
Reset of ventilator.
Description
Low Battery. Connect to Mains.
Medium
This alarm will be given when the last battery source (internal
battery) has 15 minutes of operating time left with current
settings.
The Vivo 3 will continue treatment according to the current
settings.
Connection of another power source within 15 minutes or
the treatment will shut down.
Description
Critically Low Battery. Connect to Mains
High
This alarm will be given when the last battery source (internal
battery) has 5 minutes of operating time left with current
settings.
The Vivo 3 will continue treatment according to the current
settings.
Connection of another power source within 5 minutes or the
treatment will shut down.
Doc. 007232 Y-2

Advertisement

Table of Contents
loading

Table of Contents