7.4.5

Ambient Pressure Compensation Lost Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
7.4.6

Low Battery Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset Criteria
7.4.7

Critically Low Battery Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset Criteria
66 Alarms
User Manual
Description
Pressure Comp Lost
Medium
An Ambient Pressure Compensation Lost alarm will be given
when the automatic ambient pressure compensation functionality
is out of order.
The Vivo 3 will continue treatment according to the current
settings.
Normal atmospheric pressure at sea level will be used as approxi-
mation for the temporary ambient pressure compensation. If
used at other altitude, delivered and measured pressures may
deviate.
Reset of ventilator.
Description
Low Battery. Connect to Mains.
Medium
This alarm will be given when the last battery source (internal
battery) has 15 minutes of operating time left with current
settings.
The Vivo 3 will continue treatment according to the current
settings.
Connection of another power source within 15 minutes or the
treatment will shut down.
Description
Critically Low Battery. Connect to Mains
High
This alarm will be given when the last battery source (internal
battery) has 5 minutes of operating time left with current settings.
The Vivo 3 will continue treatment according to the current
settings.
Connection of another power source within 5 minutes or the
treatment will shut down.
Doc. 007231 F-3