Ambient Pressure Compensation Lost; Alarm; High Patient Air Temp; Temperature) - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.4.5
Ambient Pressure Compensation Lost Alarm
Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
7.4.6
High Patient Air Temp. (High Patient Air Temperature)
Property
Alarm text
Priority
Alarm condition
Possible cause
Ventilator action
Reset criteria
Doc. 007229 D-3
Description
Pressure Comp Lost
Medium
An Ambient Pressure Compensation Lost alarm will be
given when the automatic ambient pressure compensation
functionality is out of order.
The Vivo 2 will continue treatment according to the current
settings.
Normal atmospheric pressure at sea level will be used as
approximation for the temporary ambient pressure compen-
sation. If used at other altitude, delivered and measured pres-
sures may deviate.
Reset of ventilator.
Description

High Patient Air Temp

High
A High Patient Air Temperature alarm will be given when
the patient air temperature exceeds 43°C (109.4°F).
If having a patient circuit with an active heated circuit, the
patient air temperature is measured by the circuit's tempera-
ture sensor. Otherwise, the temperature is measured by the
flow measurement sensor inside the Vivo 2.
Blocked air inlets.
Too high ambient temperature.
The Vivo 2 will continue treatment.
If a heated circuit or attachable humidifier is used, these will
be turned off.
The ventilator stops treatment and gives alarm for up to
2 minutes.
The temperature goes below the limit again.
Alarms
107
User manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivo 3

Table of Contents