6.4.19
Ambient Pressure Compensation Lost Alarm
Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
6.4.20
Temperature Comp. Lost (Ambient Temperature Compensation Lost Alarm)
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
120
Alarms
User manual
Description
Pressure Comp Lost
Medium
An Ambient Pressure Compensation Lost alarm will be given
when the automatic ambient pressure compensation functional-
ity is out of order.
The Vivo 65 will continue treatment according to the current
settings.
Normal atmospheric pressure at sea level will be used as
approximation for the temporary ambient pressure compensa-
tion. If used at other altitude, delivered and measured pressures
may deviate.
Reset of ventilator.
Description
Temperature Comp. Lost
An Ambient Temperature Compensation Lost alarm will be
given when the automatic ambient temperature compensation is
out of order.
Medium
The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Ambient temperature inside valid range.
Doc. 006150 M-1