Property
Priority
Ventilator action
6.4.5
Lost Mains Alarm
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.6
Crit. Low Last Power Source Alarm
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
114
Alarms
User manual
Description
Medium
The ventilator will continue treatment with the same settings.
Description
Lost Mains Power
A Mains Power Lost alarm will be given when the ventilator
switched from AC power (Mains) to another power source due
to AC Power (Mains) is lost.
Medium
The ventilator will continue treatment with the same settings.
An information message will be shown on the screen.
Confirmation by user or AC power (Mains) reconnected.
Description
Crit. Low Last Power Source
A Crit. Low Last Power Source alarm will be given when the
last battery source (internal battery or click-in battery) has
5 minutes of operating time left with current settings.
High
The ventilator will continue treatment with the same settings.
Connection of "higher" power source.
Doc. 006150 M-1