Low Minute Volume Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.6

Low Minute Volume Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Doc. 007229 D-3
Description
Low MV
High
A Low Minute Volume alarm will be given when the
minute volume is below the set alarm limit for
15 1 seconds.
Disconnection of patient circuit.
Mismatch between pressure setting and alarm setting.
Leakage from the mask or other components of the patient
circuit.
Decreased breath rate.
The minute volume goes above the alarm limit.
The Vivo 2 will continue treatment according to the cur-
rent settings.
1 l. to 30 l.
Off
0.5 l.
Alarms
97
User manual