7.3.6
Low Minute Volume Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Doc. 007229 D-3
Description
Low MV
High
A Low Minute Volume alarm will be given when the
minute volume is below the set alarm limit for
15 1 seconds.
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the patient
circuit.
•
Decreased breath rate.
The minute volume goes above the alarm limit.
The Vivo 2 will continue treatment according to the cur-
rent settings.
•
1 l. to 30 l.
•
Off
0.5 l.
Alarms
97
User manual