High Vt I (High Inspired Tidal Volume Alarm) - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.3.5
High Vt
Property
Alarm text
Priority
Alarm Condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
94
Alarms
User manual
(High Inspired Tidal Volume Alarm)
i
Description
High Vti
Medium
A High Inspired Tidal Volume alarm will be given when the
monitored Inspired Tidal Volume exceeds the set limit for the
High Inspired Tidal Volume alarm for 15 seconds.
This alarm is only used if the Vivo 65's patient circuit type is set
to Exhalation valve (single limb) circuit or in MPV modes.
Mismatch between Inspired Tidal Volume and alarm setting.
Pressure settings causing the Inspired Tidal Volume to
exceed the set alarm level.
Leakage from the mask or other components of the patient
circuit.
Mismatch between selected and used patient circuit.
When inspired tidal volume is below set alarm limit
300 ml (adult), 50 ml (paediatric) to 3000 ml (adult), 600 ml
(paediatric)
Off
10 below 600 ml, 100 above 600 ml
The ventilator will continue treatment with the same settings.
Doc. 006150 M-1