Chapter G - COLUMN
3 REPLACING THE COLUMN MOTOR
3.1 Replacing the motor
WARNING
Care must be taken when replacing the column motor. Removing parts of the lifting
assembly may enable the column to fall causing a dangerous situation and/or damage
to the equipment. The downward movement of the telescopic column must be elimi-
nated before detaching the parts.
WARNING
The X-ray unit contains live mains voltage parts. Always unplug the unit from mains
outlet before attempting to service the Input module.
CAUTION
At least two persons are needed when replacing the column motor. Never try to replace
the motor on your own.
G-12 ProMax X-ray unit with Dimax3
REPLACING THE COLUMN MOTOR
Technical Manual