Replacing Gui Software Chip, Only For Black/White Gui - Planmeca ProMax X-ray Instruction Manual

With dimax3
Hide thumbs Also See for ProMax X-ray:
Table of Contents

Advertisement

REPLACING THE GUI SOFTWARE

3.2 Replacing GUI software chip, only for black/white GUI

WARNING
NOTE
NOTE
IC 1
NOTE
Technical Manual
Always turn the X-ray unit off before removing the software chip from its socket.
Never turn the unit on, if the software chips are not in their sockets.
Antistatic precautions must be performed when handling the software chip. Touch any
grounded metal part of the unit before touching the software chip.
The GUI software update by changing the software chip can be ONLY done for a black/
white GUI.
Detach the user interface according to the instructions given in section 2.2 "Support arm
lower cover" on page F-8.
Remove the software chip from its socket marked IC16 with the special tool.
FLASH 3
FLASH 4
Figure 14
If an IC1 chip is present then remove it also.
Carefully place the new software chip into the socket.
Attach the user interface to the patient support arm
IC 16
+
GUI
Chapter F - PATIENT SUPPORT ARM
FLASH 2
FLASH 1
ProMax X-ray unit with Dimax3 F-11

Advertisement

Table of Contents
loading

Table of Contents