ADJUSTMENTS AND CALIBRATIONS
Technical manual
The Ceph mode beam position display appears. The sensor head and the first primary colli-
mator will move to the centre position. Touch the Test field.
Figure 47
Place the beam alignment tool to the sensor alignment tool to lower vertical position as shown
on the Fig. 48 below.
Beam alignment
tool in upper
position
Figure 48
Stand behind the tube head and protect yourself from radiation. Press and hold down the
exposure button. The image of the radiation beam will appear on the beam alignment tool.
Chapter H - CEPHALOSTAT
Beam alignment
tool in lower
position
ProMax X-ray unit with Dimax3 H-27