ADJUSTMENTS AND CALIBRATIONS
Chapter H - CEPHALOSTAT
Select the vertical asymmetric image orientation as follows: Touch the Image size field on the
Main display. Select image size & orientation display appears. Select the vertical asymmetric
image orientation by touching the respective image icon on the display.
Touch the i-field on the Main display. The list of Information displays appears.
Figure 86
From the list of Information displays select Technical calibrations (i600) .
The Technical calibrations mode is password protected and the password is asked when the
mode is entered for the first time after switching the unit on. The password is 1701.
Figure 87
ProMax X-ray unit with Dimax3 H-47