HELP AND ERROR MESSAGE SHORT-FORM TABLES
Error
Explanation
code
E921
Program
errors
E922
E923
E924
E925
E926
E927
E928
E929
E931
Hardware errors
Technical Manual
structure
Stack overflow
Out of RAM
Watchdog reset
Unimplemented opcode trap
Illegal program vector
Unexpected reset
PathSolver out of memory
PathSolver cannot solve trajectory
DSP program error (reserved)
RTC-clock runs or CPU clock is running
at wrong speed
Chapter C - ERROR MESSAGES
Comments
Restart thew ProMax X-ray unit.
Switch off the X-ray unit and
update the ProMax X-ray unit
software.
Switch off the X-ray unit and
update the ProMax X-ray unit
software.
Switch off the X-ray unit and
update the ProMax X-ray unit
software.
Switch off the X-ray unit and
update the ProMax X-ray unit
software.
Restart the ProMax X-ray unit.
Recalibrate the layer lights again.
The movement of the ProMax X-
ray unit arm is not possible.
Switch off the ProMax X-ray unit
and move the arms to a new safe
position. Perform the Configura-
tion selftest (i560) and the Posi-
tion sensor calibration (i640)
and restart the X-ray unit.
Check the functionality of the
RTC PCB, the cable and the bat-
tery.