Chapter C - ERROR MESSAGES

1.10 Infrastructure errors (9xx)

Error
Explanation
code
E901
Checksum error in
E902
E903
E904
E905
E906
E907
E908
E911
RAM errors
E912
E913
E914
C-16 ProMax X-ray unit with Dimax3
Main program FLASH, Low Block
Main program FLASH, High Block
Memory Expansion module FLASH
RTC PCB memory problem
Tube head FBK PCB memory error
PAN AEC memory problem
CEPH AEC memory problem
Autofocus module problem (reserved)
DATA RAM BIT ERROR (even bank)
DATA RAM BIT ERROR (odd bank)
FPGA register RAM error
CPU internal RAM error
HELP AND ERROR MESSAGE SHORT-FORM TABLES
Comments
Restart the ProMax X-ray unit.
Update the ProMax X-ray unit
software version.
Update the ProMax X-ray unit
software version.
Check the RTC PCB cabling and
the battery.
Check the cabling and the func-
tionality of the FBK PCB.
Change the cable / the PSU
PCB.
Change the cable / the PSU
PCB.
Check the CPU PCB and the
cabling.
Check the CPU PCB and the
cabling.
Check the CPU PCB and the
cabling.
Check the CPU PCB and the
cabling.