Chapter A - GENERAL & TECHNICAL DATA
Guidance and manufacturer's declaration - electromagnetic immunity
Planmeca ProMax X-ray unit is intended for use in the electromagnetic environment specified below. The
customer or the user of Planmeca ProMax X-ray unit should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency(
50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
A-14 ProMax X-ray unit with Dimax3
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line to line
±2 kV line to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line to line
±2 kV line to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
EMC INFORMATION
Electromagnetic environment-
guidance
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of Planmeca ProMax X-ray unit
requires continued operation dur-
ing power mains interruptions, it
is recommended that Planmeca
ProMax X-ray unit be powered
from an uninterruptible power
supply.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment. The power frequency mag-
netic field should be measured in
the intended installation location
to assure that it is sufficiently low.
Technical Manual