ADJUSTMENTS AND CALIBRATIONS
Chapter H - CEPHALOSTAT
To adjust the collimator plate position sideways tighten one angle adjustment screw as much
as you loosen the other.
Angle adjustment screws
Figure 54
Check the adjustments and tighten the attachment screws.
Adjusting the vertical position of the second primary collimator
Loosen the two vertical adjustment locking screws with 3mm Allen key.
Collimator vertical position locking screw
Figure 55
ProMax X-ray unit with Dimax3 H-31
Technical manual