Chapter B - INFORMATION DISPLAYS
5 SERVICE SETTINGS
From the list of information displays select Service settings (i500) . The service setting mode is
password protected and the password is asked when the Service settings mode is entered for
the first time after switching the unit on. The password is 1701 .
Figure 33
In Service settings mode the Special operation mode selection (i510) and exposure program
tuning (i540) can be made, Preheat calibration (i520) can be performed, Maximum height
(i530) of the unit can be set and Exposure program tuning (i540) can be made.
Also the factory Default settings (i550) can be recalled and Configuration selftest (i560) can
be made. This mode allows to perform the fine tuning of the Patient support alignment (i580).
The GUI software is updated by using the Software update (i590) menu.
Figure 34
B-22 ProMax X-ray unit with Dimax3
SERVICE SETTINGS
Technical Manual