ADJUSTMENTS
Chapter F - PATIENT SUPPORT ARM
Move the thumb wheel of the layer light slightly to switch the three patient positioning lights
on. The layer light thumb wheel is located on the underside of the patient support table. Move
the layer light beam so that it is on the black reference line on the side of the ball phantom.
Accept the new the layer light zero position and exit the calibration mode by touching the
Done field.
Ball phantom
Layer light beam
Figure 8
If the light beam is not vertical, or it is in focus, it must be adjusted according to the instruc-
tions in section "Layer light adjustment" on page F-6.
ProMax X-ray unit with Dimax3 F-5
Technical Manual