ADJUSTMENTS AND CALIBRATIONS
Chapter H - CEPHALOSTAT
From the list on the i600 display that appears select Primary collimator calibration (i610) and
from the list on the i610 display that appears select the Ceph mode beam position (i616) .
Figure 20
The Ceph mode beam position display appears. The sensor head and the first primary colli-
mator will move to the centre position. Touch the Test field.
Figure 21
ProMax X-ray unit with Dimax3 H-11
Technical manual