Adjusting The Telescopic Column Position - Planmeca ProMax X-ray Instruction Manual

With dimax3
Hide thumbs Also See for ProMax X-ray:
Table of Contents

Advertisement

ADJUSTMENTS

1.2 Adjusting the telescopic column position

Loosen left side
attachment screws.
Technical Manual
With the height adjusting buttons drive the unit to the position shown on Fig. 3 below.
Height adjusting buttons
With a spirit level make sure that the telescopic column is parallel with the stationary column
both in depth and sideways. To adjust the sideways position loosen the two attachment
screws on the left side of the column. Rotate the nuts located between the columns so that
the space between the columns on the left side is equal to the space on the right side, and the
telescopic column is parallel with the stationary column. Tighten the upper attachment screw.
Figure 3
Down
Adjust the column sideways
position with these nuts.
Chapter G - COLUMN
Up
ProMax X-ray unit with Dimax3 G-3