Operating the TSG-3800
Setting Up Input Reference Controls
Table 3-2. MTIE Pre-defined Settings (Continued)
Configuration
MCD
Table 3-3. Frequency Pre-defined Settings
Configuration
PRS
OCN
(Factory Default Setting)
MCD
For Frequency and MTIE measurements, you can set a threshold and fault mode for
each of three intervals (100, 1000, or 10000 seconds). If you do not define these
limits, they remain at the default values; see
108. The MTIE and Frequency command syntax is described in
on page 352, and
For example, you can assign a REPORT fault mode for a fractional frequency offset
exceeding 1 x 10E
can specify an offset threshold for each time scale (T100, T1000, or T10000) for
both limits.
Note: IGNORE or ALARM fault modes are the other settings for L1
and L2. For consistency, use L1 for the more extreme limit and fault
mode.
92
TSG-3800 User's Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Time Scale
T100
T1000
T10000
Time Scale
T100
T1000
T10000
T100
T1000
T10000
T100
T1000
T10000
Frequency
Command, on page 348.
–8
and a FAIL fault mode for an offset exceeding 1 x 10E
Limit 1 (Fail)
Limit 2 (Alarm)
432
324
864
648
1728
1296
Limit 1 (Fail)
Limit 2 (Alarm)
5000
2000
500
200
50
20
5000
2000
750
300
60
30
8000
4000
800
400
80
40
Default Parameter
Values, on page
MTIE
12778474-002-2 Revision F – March 2004
Command,
–7
. You