The Retrieve PM Frequency command reports the frequency data stored in the 1-s,
100-s, or 1000-s frequency history buffers for the specified port. You can specify the
number of data points to display; the TSG-3800 stores up to 50 data points.
If the TSG-3800 successfully receives the command, it returns the COMPLD
message. If the TSG-3800 does not successfully receive the command, it returns
the DENY message.
Command Syntax
Valid aid
Minimum User Level
Response Format
Example Command
Example Response
404 TSG-3800 User's Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
RTRV-PM-FREQ:[tid]:[aid]:ctag[::tmper]
[:points]];
where:
tmper is 1, 100, or 1000
points is 1–50. Default is 1
INPx[-y]
1
<cr><lf><lf>
M^ctag^COMPLD<cr><lf>
^^^"aid:ocrdat,ocrtim,FREQ,tmper,monval[<cr><
lf>;
where:
ocrdat and ocrtime are the actual date and time the data
were collected
FREQ describes the data that follows
tmper is the data measurement time period in seconds
monval is the frequency value for the specified input port
in ps/s with commas separating the values
RTRV-PM-FREQ:TSG:INP3-3:ctag;
TSG 02-09-10 19:00:19
M 13 COMPLD
^^^
;