5
Commands by Subsystem
:TIMebase:WINDow:POSition
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
The :TIMebase:WINDow:POSition command sets the horizontal position in
the zoomed (delayed) view of the main sweep. The main sweep range and
the main sweep horizontal position determine the range for this command.
The value for this command must keep the zoomed view window within
the main sweep range.
Query Syntax
:TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal
window position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also
•
•
•
•
•
•
•
408
(see
page
658)
view reference point in NR3 format
"Introduction to :TIMebase Commands"
":TIMebase:MODE"
on page 402
":TIMebase:POSition"
":TIMebase:RANGe"
on page 404
":TIMebase:SCALe"
on page 406
":TIMebase:WINDow:RANGe"
":TIMebase:WINDow:SCALe"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 400
on page 403
on page 409
on page 410