Trigger[:Edge]:Level - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:TRIGger[:EDGE]:LEVel

Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
<level> ::= ±(external range setting) in NR3 format
<source> ::= {CHANnel<n> | EXTernal}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the
active trigger source.
If the optional source is specified and is not the active source, the level on the active source
N O T E
is not affected and the active source is not changed.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current
trigger source.
Return Format
<level><NL>
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
for internal triggers
for external triggers
"Introduction to :TRIGger Commands"
":TRIGger[:EDGE]:SOURce"
":EXTernal:RANGe"
on page 226
on page 411
on page 444
Commands by Subsystem
5
441

Hide quick links:

Advertisement

Table of Contents
loading