Waveform:byteorder - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:WAVeform:BYTeorder

Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD
data. The parameter MSBFirst sets the most significant byte to be
transmitted first. The parameter LSBFirst sets the least significant byte to
be transmitted first. This command affects the transmitting sequence only
when :WAVeform:FORMat WORD is selected. The default setting is
LSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also
Example Code
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
"Introduction to :WAVeform Commands"
":WAVeform:DATA"
on page 509
":WAVeform:FORMat"
":WAVeform:PREamble"
"Example Code"
on page 521
"Example Code"
on page 517
on page 502
on page 511
on page 516
Commands by Subsystem
5
507

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents