:TIMebase:WINDow:POSition
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the delayed view
The :TIMebase:WINDow:POSition command sets the horizontal position in
the delayed view of the main sweep. The main sweep range and the main
sweep horizontal position determine the range for this command. The
value for this command must keep the delayed view window within the
main sweep range.
Query Syntax
:TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal
window position setting in the delayed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also
•
•
•
•
•
•
•
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
reference point in NR3 format
"Introduction to :TIMebase Commands"
":TIMebase:MODE"
on page 345
":TIMebase:POSition"
":TIMebase:RANGe"
on page 347
":TIMebase:SCALe"
on page 349
":TIMebase:WINDow:RANGe"
":TIMebase:WINDow:SCALe"
on page 343
on page 346
on page 352
on page 353
Commands by Subsystem
5
351