Trigger:lin:id - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:TRIGger:LIN:ID

Command Syntax
:TRIGger:LIN:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :TRIGger:LIN:ID command defines the LIN identifier searched for in
each CAN message when the LIN trigger mode is set to frame ID.
This command is only valid when the automotive CAN and LIN serial decode option (Option
N O T E
AMS) has been licensed.
Setting the ID to a value of "- 1" results in "0xXX" which is equivalent to
all IDs.
Query Syntax
:TRIGger:LIN:ID?
The :TRIGger:LIN:ID? query returns the current LIN identifier setting.
Return Format
<value><NL>
<value> ::= integer in decimal
Errors
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
from 0-63 or 0x00-0x3f
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 417
":TRIGger:LIN:TRIGger"
":TRIGger:LIN:SIGNal:DEFinition"
":TRIGger:LIN:SOURce"
on page 617
on page 411
on page 469
on page 612
on page 466
Commands by Subsystem
5
463