Measure:define - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

5
Commands by Subsystem

:MEASure:DEFine

Command Syntax
:MEASure:DEFine <meas_spec>
<meas_spec> ::= {DELay | THResholds}
The :MEASure:DEFine command sets up the definition for measurements
by specifying the delta time or threshold values. Changing these values
may affect the results of other measure commands. The table below
identifies which measurement results that can be affected by redefining
the DELay specification or the THResholds values. For example, changing
the THResholds definition from the default 10%, 50%, and 90% values may
change the returned measurement result.
MEASure Command
DUTYcycle
DELay
FALLtime
FREQuency
NWIDth
OVERshoot
PERiod
PHASe
PREShoot
PWIDth
RISetime
VAVerage
VRMS
:MEASure:DEFine
:MEASure:DEFine DELay,<delay spec>
DELay Command
<delay spec> ::= <edge_spec1>,<edge_spec2>
Syntax
<edge_spec1> ::= [<slope>]<occurrence>
<edge_spec2> ::= [<slope>]<occurrence>
<slope> ::= {+ | -}
<occurrence> ::= integer
276
(see
page
658)
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
DELay
x
THResholds
x
x
x
x
x
x
x
x
x
x
x
x
x

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents