Agilent Technologies InfiniiVision 5000 Series Programmer's Manual page 278

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

5
Commands by Subsystem
Return Format
for <meas_spec> = DELay:
{ <edge_spec1> | <edge_spec2> | <edge_spec1>,<edge_spec2>} <NL>
for <meas_spec> = THResholds and <threshold mode> = PERCent:
THR,PERC,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
for <meas_spec> = THResholds and <threshold mode> = ABSolute:
THR,ABS,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
for <threshold spec> = STANdard:
THR,PERC,+90.0,+50.0,+10.0
See Also
278
"Introduction to :MEASure Commands"
":MEASure:DELay"
on page 279
":MEASure:SOURce"
":CHANnel<n>:RANGe"
":CHANnel<n>:SCALe"
":CHANnel<n>:PROBe"
":CHANnel<n>:UNITs"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
and lower threshold percentage values
between Vbase and Vtop in NR3 format.
and lower threshold voltages in NR3
format.
on page 272
on page 297
on page 204
on page 205
on page 199
on page 206

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents