:MTEenable (Mask Test Event Enable Register)

Command Syntax
:MTEenable <mask>
<mask> ::= 16-bit integer
The :MTEenable command sets a mask in the Mask Test Event Enable
register. Set any of the following bits to "1" to enable bit 9 in the
Operation Status Condition Register and potentially cause an SRQ (Service
Request interrupt to be generated.
Table 38 Mask Test Event Enable Register (MTEenable)
Bit
15-11
10
9
8
7-2
1
0
Query Syntax
:MTEenable?
The :MTEenable? query returns the current value contained in the Mask
Test Event Enable register as an integer number.
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
Name
Description
---
---
Auto
Auto Mask Created
Mask
---
---
Started
Mask Testing Started
---
---
Fail
Mask Test Fail
Comple
Mask Test Complete
te
Commands by Subsystem
When Set (1 = High = True), Enables:
(Not used.)
Auto mask creation completed.
(Not used.)
Mask testing started.
(Not used.)
Mask test failed.
Mask test is complete.