5
Commands by Subsystem

:MTESt:SCALe:BIND

Command Syntax
:MTESt:SCALe:BIND <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:SCALe:BIND command enables or disables Bind 1 & 0 Levels
(Bind - 1 & 0 Levels for inverted masks) control:
• ON —
• OFF —
Query Syntax
:MTESt:SCALe:BIND?
The :MTESt:SCALe:BIND? query returns the value of the Bind 1&0 control
(Bind - 1&0 for inverted masks).
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
344
(see
page
658)
If the Bind 1 & 0 Levels control is enabled, the 1 Level and the 0 Level
controls track each other. Adjusting either the 1 Level or the 0 Level
control shifts the position of the mask up or down without changing its
size.
If the Bind - 1 & 0 Levels control is enabled, the - 1 Level and the 0
Level controls track each other. Adjusting either the - 1 Level or the 0
Level control shifts the position of the mask up or down without
changing its size.
If the Bind 1 & 0 Levels control is disabled, adjusting either the 1 Level
or the 0 Level control changes the vertical height of the mask.
If the Bind - 1 & 0 Levels control is disabled, adjusting either the - 1
Level or the 0 Level control changes the vertical height of the mask.
"Introduction to :MTESt Commands"
":MTESt:SCALe:X1"
on page 345
":MTESt:SCALe:XDELta"
":MTESt:SCALe:Y1"
on page 347
":MTESt:SCALe:Y2"
on page 348