:MTESt:RMODe:SIGMa

Command Syntax
:MTESt:RMODe:SIGMa <level>
<level> ::= from 0.1 to 9.3 in NR3 format
When the :MTESt:RMODe command is set to SIGMa, the
:MTESt:RMODe:SIGMa command sets the test sigma level to which a mask
test runs. Test sigma is the best achievable process sigma, assuming no
failures. (Process sigma is calculated using the number of failures per
test.) The test sigma level indirectly specifies the number of waveforms
that must be tested (in order to reach the sigma level).
Query Syntax
:MTESt:RMODe:SIGMa?
The :MTESt:RMODe:SIGMa? query returns the current Sigma level setting.
Return Format
<level><NL>
<level> ::= from 0.1 to 9.3 in NR3 format